Monitoring biomolecule concentrations in tissue using a wearable droplet microfluidic-based sensor

Adrian M Nightingale, Chi Leng Leong, Rachel A Burnish, Sammer-Ul Hassan, Yu Zhang, Geraldine F Clough, Martyn G Boutelle, David Voegeli, Xize Niu

Research output: Contribution to journalArticlepeer-review

Abstract

Knowing how biomarker levels vary within biological fluids over time can produce valuable insight into tissue physiology and pathology, and could inform personalised clinical treatment. We describe here a wearable sensor for monitoring biomolecule levels that combines continuous fluid sampling with in situ analysis using wet-chemical assays (with the specific assay interchangeable depending on the target biomolecule). The microfluidic device employs a droplet flow regime to maximise the temporal response of the device, using a screw-driven push-pull peristaltic micropump to robustly produce nanolitre-sized droplets. The fully integrated sensor is contained within a small (palm-sized) footprint, is fully autonomous, and features high measurement frequency (a measurement every few seconds) meaning deviations from steady-state levels are quickly detected. We demonstrate how the sensor can track perturbed glucose and lactate levels in dermal tissue with results in close agreement with standard off-line analysis and consistent with changes in peripheral blood levels.
Original languageEnglish
JournalNature Communications
Volume10
Issue number1
DOIs
Publication statusPublished - 21 Jun 2019

Cite this